JF-5A Digital Display Temperature Controlled Oxygen Index Tester
Technical parameter
Environmental temperature: 23 ºC± 2 ºC
Relative humidity:< 85%
Voltage and power: 220V, 50HZ, 100W
Gas: GB3868 Industrial Gaseous Oxygen; GB3864 Industrial gaseous oxygen; Both bottles of gas require pressure regulating valves.
Input pressure: 0.2-0.3MPa
Weight: Approximately 25Kg
Working pressure: Nitrogen 0.1-0.15MPa Oxygen 0.1-0.15Mpa
Testing item
It is suitable for the combustion performance test of wood, plastic, rubber, fiber, foam plastic and other materials.
Applicable standard
GB/T2406.2 GB/T5454 GB/T10707 ASTM D2863IS0 4589-2
1. Adopting imported oxygen sensors, digital displays oxygen concentration ( not need to calculate), also with higher and more accurate precision and a range of 0-100%.
2. Digital resolution:+0.1%
3. Measurement accuracy: 0.4 level
4. Flow regulation range: 0-10L/min (60-600L/h)
5. Response time:< 5s
6. Combustion cylinder: inner diameter>75mm, height 400mm
7. Gas flow velocity inside the combustion cylinder: 40mm+10mm/s
8. Pressure gauge accuracy level: 2.5, resolution: 0.01MPa
9. Flow meter:1-15L/min (60-900L/H) adjustable
JF-5 Digital Display Oxygen Index Tester
Applicable standard
The JF-5 oxygen index analyzer is a new type of product developed according to the technical requirements specified in the national standard GB/T2406.
Technical parameter
1. Adopting imported oxygen sensors, digital displays oxygen concentration ( not need to calculate), also with higher and more accurate precision and a range of 0-100%.
2. Digital resolution:+0.1%
3. Measurement accuracy: 0.4 level
4. Flow regulation range: 0-10L/min (60-600L/h)
5. Response time:< 5s
6. Combustion cylinder: inner diameter>75mm, height 400mm
7. Gas flow velocity inside the combustion cylinder: 40mm+10mm/s
8. Pressure gauge accuracy level: 2.5, resolution: 0.01MPa
9. Flow meter:1-15L/min (60-900L/H) adjustable